Accession Number : AD0601561

Title :   A METHOD OF MEASURING TEMPERATURE WITH A PLATINUM RESISTANCE THERMOMETER,

Corporate Author : FOREIGN TECHNOLOGY DIV WRIGHT-PATTERSON AFB OHIO

Personal Author(s) : Alieva,F. Z.

Report Date : 26 MAY 1964

Pagination or Media Count : 5

Abstract : The method described simplifies the process of temperature determination. The ratio of the thermometer resistance at 100 degrees to the resistance at 0 degrees is established by calibration of the platinum resistance thermometer. The ratio is reduced to the standard magnitude 1.392. The measured temperature can then be calculated from the known formula given. 16

Descriptors :   (*TEMPERATURE, MEASUREMENT), (*RESISTANCE THERMOMETERS, PLATINUM), CALIBRATION

Distribution Statement : APPROVED FOR PUBLIC RELEASE