Accession Number : AD0622778

Title :   MODIFICATION OF TEST SET, CRYSTAL UNIT, QUARTZ TS-710/TSM.

Descriptive Note : Technical rept.,

Corporate Author : ARMY ELECTRONICS COMMAND FORT MONMOUTH N J

Personal Author(s) : Shibla,Charles L.

Report Date : JUL 1965

Pagination or Media Count : 40

Abstract : A modification to Test Set, Crystal Unit, Quartz TS710/TSM is described. The modification is comprised of revisions in several circuit areas with a resultant improvement in operation. Overdrive of low-resistance crystals is minimized, crystal power is more accurately and conveniently measured, and frequency/resistance calibration with any standard military crystal unit is possible. (Author)

Descriptors :   (*TEST SETS, CRYSTALS), QUARTZ, PERFORMANCE(ENGINEERING), LOW FREQUENCY, VERY LOW FREQUENCY, MEDIUM FREQUENCY

Distribution Statement : APPROVED FOR PUBLIC RELEASE