Accession Number : AD0640318

Title :   MOSSBAUER-EFFECT SPECTROSCOPY,

Corporate Author : ARMY NUCLEAR DEFENSE LAB EDGEWOOD ARSENAL MD

Personal Author(s) : Temperley,A. A.

Report Date : OCT 1966

Pagination or Media Count : 36

Abstract : Mossbauer-effect spectroscopy is discussed with emphasis on spectrometer design. Construction of a Mossbauer-effect spectrometer using a loudspeaker as the velocity transducer, capable of operation with either constant velocities or with constant accelerations, is described. Resonance experiments with the 14.4-keV Mossbauer-effect gamma ray from Co57 (Fe57m) are used to describe the performance of the spectrometer. (Author)

Descriptors :   (*MOSSBAUER EFFECT, SPECTROSCOPY), (*NUCLEAR RADIATION SPECTROSCOPY, MOSSBAUER EFFECT), INSTRUMENTATION, NUCLEAR RADIATION SPECTROMETERS, TRANSDUCERS, COBALT, GAMMA EMISSION, IRON, RADIOACTIVE ISOTOPES

Subject Categories : Test Facilities, Equipment and Methods

Distribution Statement : APPROVED FOR PUBLIC RELEASE