Accession Number : AD0646024

Title :   HIGH INTENSITY VUV ARC SOURCE,

Corporate Author : VITRO LABS WEST ORANGE N J WEST ORANGE LAB

Personal Author(s) : Levy,M. E. ; Huffman,R. E.

Report Date : 01 APR 1966

Pagination or Media Count : 25

Abstract : A novel short arc configuration was operated in argon and helium. Vacuum ultraviolet radiation in the spectral range 2000-200 A was observed through a port in the anode. The arc was operated windowless at pressures up to 2000 Torr in conjunction with a three stage differentially pumped vacuum system. A Seya-Namioka half meter monochromator was used for the observations with a 1200 1/mm grating blazed for 1500 A and coated with MgF2. A sodium salicylate detector and an aluminum-terphenyl filter detector combination were employed. The radiating plasma was found to be at LTE with an electron density of 5. 10 to the 16th power/cu cm and a temperature of 1.5 ev. (Author)

Descriptors :   (*ULTRAVIOLET RADIATION, VACUUM APPARATUS), ELECTRIC ARCS, SOURCES, ULTRAVIOLET SPECTROSCOPY, PLASMAS(PHYSICS), ARGON, HELIUM

Subject Categories : Optics