Accession Number : AD0649231

Title :   ON THE ACCURACY OF MEASUREMENTS WITH QUARTZ GRAVIMETERS WITH A HORIZONTAL TORSIONAL THREAD,

Corporate Author : FOREIGN TECHNOLOGY DIV WRIGHT-PATTERSON AFB OHIO

Personal Author(s) : Bulanzhe,Yu. D.

Report Date : JAN 1960

Pagination or Media Count : 26

Abstract : It was established that in quartz gravimeters with a horizontal torsional thread and liquid temperature compensation (Norgaard, SN-3, and GAE-2 gravimeters) the error of the reading is little affected by the measured angle of spread, and it is generally characterized by a value of plus or minus 0.20 to 0.30 mgal. (Author)

Descriptors :   (*MEASURING INSTRUMENTS, GRAVITY), ACCURACY, QUARTZ, ERRORS, USSR