Accession Number : AD0649809

Title :   THERMAL FLUX DISTRIBUTION IN AN ARC-IMAGING FURNACE.

Descriptive Note : Technical rept.,

Corporate Author : ARMY MATERIALS RESEARCH AGENCY WATERTOWN MASS

Personal Author(s) : Farrow,Raymond L. ; Matkovich,Michael J.

Report Date : DEC 1966

Pagination or Media Count : 22

Abstract : To effectively utilize an arc-imaging furnace as a source of controlled thermal energy, the flux density received at the sample must be known. A commercially available, double-elliptical, arc-imaging furnace has been calibrated using both an integrating calorimeter and a discriminating radiometer. The results show the thermal flux received over both a 0.9-mm-diameter area and a 9.5-mm-diameter area as a function of three-dimensional position within the receiving zone. In addition, flux contour maps are presented for several appropriate planes in the focal zone. The effects of rotating interrupter shutters on both heat flux and optical temperature measurements are discussed.

Descriptors :   (*THERMAL RADIATION, *CALORIMETERS), (*LABORATORY FURNACES, RADIOMETERS), HEAT FLUX, HEAT TRANSFER, HIGH TEMPERATURE, PYROMETERS, OPTICAL INSTRUMENTS, EMISSIVITY