Accession Number : AD0652481

Title :   FABRICATION OF LITHIUM-DRIFT GERMANIUM RADIATION DETECTORS.

Descriptive Note : Master's thesis,

Corporate Author : AIR FORCE INST OF TECH WRIGHT-PATTERSON AFB OHIO SCHOOL OF ENGINEERING

Personal Author(s) : Mullady,James Benedict

Report Date : JUN 1966

Pagination or Media Count : 55

Abstract : A simple technique for fabricating lithium-drift germanium radiation detectors for use in gamma-ray spectroscopy is described in detail. Lithium-ion drift is conducted in room atmosphere at constant temperature and voltage for 470 hours to obtain a 2 mm thick sensitive region 7.5cm2 in area. One detector made by this method gives 10.3 kev (fwhm) resolution for the Cs137 662 kev gamma-ray. Leakage current at 77K and 65 volts reverse bias is 1.65 x 0.0000001 A.

Descriptors :   (*GAMMA RAY SPECTROSCOPY, DETECTORS), (*GERMANIUM, CRYSTALS), LITHIUM, IONS, TABLES(DATA), INTERACTIONS, RADIATION MEASURING INSTRUMENTS, CESIUM, SEMICONDUCTORS, VOLTAGE, LEAKAGE(ELECTRICAL)

Subject Categories : Nuclear Instrumentation
      Solid State Physics