Accession Number : AD0725502

Title :   Use of Miniature Glass-Needle-Type TL Dosimeters in Finger-Ring Applications.

Descriptive Note : Interim rept.,

Corporate Author : NAVAL RESEARCH LAB WASHINGTON D C

Personal Author(s) : Booth,Leroy F. ; Johnson,Tommy L. ; Attix,Frank H.

Report Date : 24 MAY 1971

Pagination or Media Count : 21

Abstract : Miniature glass-encapsulated needle-shaped thermoluminescent (TL) dosimeters, 1.4 mm in outer diameter and 12 mm in length, have been investigated for routine application in finger rings to monitor the dose received by the hands, e.g., in certain Linac repair operations. Both CaF2:Mn and LiF-filled dosimeters were studied, and special attention was given to the simplification or elimination of the cumbersome annealing procedures normally used with LiF dosimeters. (Author)

Descriptors :   (*DOSIMETERS, THERMOLUMINESCENCE), (*DOSE RATE, HANDS), GAMMA RAYS, RADIATION HAZARDS, ANNEALING, TEST METHODS, CURVE FITTING

Subject Categories : Radiobiology
      Nuclear Instrumentation