Accession Number : AD0863324

Title :   A Comparison between NIO/4777 Thermostat Salinity Meter Measurements and Three Chemical Analyses.

Descriptive Note : Technical memo.,

Corporate Author : SACLANT ASW RESEARCH CENTRE LA SPEZIA (ITALY)

Personal Author(s) : Dahme, A.

Report Date : 15 APR 1965

Pagination or Media Count : 13

Abstract : The SACLANT ASW Research Centre research vessel is equipped with an NIO/4777 Thermostat Salinity Meter to measure the conductivity, and hence the salinity, of sea-water specimens. The conversion from conductivity measurements to salinity measurements is by means of tables computed by the N.I.O. The tests described here were made to verify the salinity measurements obtained by means of the instrument and its conversion tables against measurements of the same water specimens made by normal chemical analysis.

Descriptors :   (*SALINOMETERS, THERMOSTATS), CHEMICAL ANALYSIS, ELECTRICAL CONDUCTIVITY, SALINITY, SEA WATER, VOLUMETRIC ANALYSIS, COMPATIBILITY, REGRESSION ANALYSIS, EQUATIONS, ERRORS, ITALY, MEDITERRANEAN SEA, OCEANOGRAPHY.

Subject Categories : Physical Chemistry
      Physical and Dynamic Oceanography
      Test Facilities, Equipment and Methods

Distribution Statement : APPROVED FOR PUBLIC RELEASE