Accession Number : ADA141288

Title :   An Integrating Dosimeter for Pulsed Radiation,

Corporate Author : ROME AIR DEVELOPMENT CENTER GRIFFISS AFB NY

Personal Author(s) : Cappelli,J R

PDF Url : ADA141288

Report Date : Dec 1983

Pagination or Media Count : 16

Abstract : A Pulsed Radiation Dosimetry System designed to measure radiation produced by particle accelerators is described. The problems associated with total-dose measurement of irregular shaped pulses of ionizing radiation have been simplified. The system responds to extremely narrow pulses of charged particles or X-rays using a pin diode as the detector. Direct readout of dose in rads (Si) is displayed on a 3-1/2 digit digital panel meter. The system will operate in either the multiple-pulse or single-pulse mode. The multiple-pulse mode would be useful in monitoring or tuning a linear accelerator. In the single-pulse mode the system will automatically display total dose of a transient event in real time and hold that measurement indefinitely or until the next measurement is made. The system features an automatic reset in either mode. (Author)

Descriptors :   *Dosimetry, *Radiation measuring instruments, *Particle accelerators, Charged particles, X rays, Pin diodes, Ionizing radiation, Circuits, Detectors, Attenuators, Integrators, Multivibrators, Monitoring

Subject Categories : Radiobiology
      Particle Accelerators

Distribution Statement : APPROVED FOR PUBLIC RELEASE