Title :   Fourier Transform Ion Cyclotron Resonance Mass Spectrometry at Low Magnetic Fields: Several Figures of Merit.

Descriptive Note : Technical rept. Jan 85-Sep 86,

Corporate Author : FLORIDA UNIV GAINESVILLE DEPT OF CHEMISTRY

Personal Author(s) : Hearn,Bryan A ; Baykut,Gokhan ; Eyler,John R

PDF Url : ADA183650

Report Date : 29 Jul 1987

Pagination or Media Count : 20

Abstract : A Fourier transform ion cyclotron resonance mass spectrometer based on a permanent magnet with a magnetic induction of 0.16 T with low homogeneity provided a precision of 3.2% (taken as the relative standard deviation of the hourly mean of three independent measurements) over a 5 hour period and a precision of 1 1.7% for daily measurements over a 10 day period. For the permanent magnet system, resolution values for perdeuterated benzene (m z 84) for a given mass range and a given number of data points were similar to those measured for both an electromagnet (0.2 to 1.0 T) and a superconducting solenoid (3.0 T). Signal to noise ratios for perdeuterated benzene obtained with the three different magnets were also measured. Keywords: Fourier Transform Ion Cyclotron Resonance; Mass Spectrometry; Instrumentation; Miniaturization.

Descriptors :   *MASS SPECTROMETRY, *MASS SPECTROMETERS, BENZENE, FIGURE OF MERIT, MINIATURIZATION, PERMANENT MAGNETS, STANDARD DEVIATION, SIGNAL TO NOISE RATIO, MEASUREMENT, HOMOGENEITY, MAGNETS, DATA BASES, RESOLUTION, SOLENOIDS, SUPERCONDUCTORS, FOURIER TRANSFORMATION, CYCLOTRON RESONANCE, DEUTERIUM COMPOUNDS, STABILITY